Error: "Vial flow not ready" and "inconsistent vial numbers" with Agilent Headspace 7697A and Empower - WKB8837

Article number: 8837

SYMPTOMS

  • 7697A picks up the vial and moves it to an inject position. Then the following error appears: "vial flow not ready" and returns the vial to its original position.
  • On the Headspace panel, the following error message appears: "external device not ready".
  • In the Empower message center, "inconsistent vial numbers". 

ENVIRONMENT

  • Agilent 7890B
  • Agilent 7697A Headspace
  • Empower 3 Service Release 2
  • Headspace Control Option Version 3.0

CAUSE

Tray Type and "start vial" and "end vial" values are incorrect.

FIX or WORKAROUND

  1. Open the Empower Instrument method for 7697A.
  2. On General tab, select the Tray type for HS 7697A (see screenshot).
  3. On Condition tab, select the Vial Start and End Vial values as shown in second screenshot. 

 

 

7697A_ General Tab.PNG

7697A_ Vial start and end.PNG

 

 

ADDITIONAL INFORMATION

The Start Vial and End Vial positions need to match what is in the Sample Set method and Empower Instrument method. 

id8837, 7890 GC, 7890 GCDET, 7890BGC, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, HEADSPACE, SUP