Agilent 6890 with G1888 Headspace intemittently fails to complete sample set - WKB87830
Article number: 87830
SYMPTOMS
- Run stops; "Preparing For Next Injection" displayed
- "Connection to Chromatographic System Lost" eventually appears
- Sample sets were altered with Alter Running Sample Set
ENVIRONMENT
- Agilent 6890 GC
- G1888
- Empower 3
- ICF 3.1
CAUSE
Samples sets cannot be edited if "High Throughput Is Enabled" is in the instrument method.
FIX or WORKAROUND
Disable "High Throughput" or do not use "Alter Running Sample".
ADDITIONAL INFORMATION
This limitation is stated in the ICF 3.1 release notes.