What are the headspace rules that apply for high throughput and low throughput mode under the Options tab? - WKB86822
Article number: 86822
ENVIRONMENT
- Agilent Instrument Control Framework
- Agilent 7697A Headspace
- Agilent G1888 Headspace
- Empower 3 Enterprise
ANSWER
When running samples, the following headspace rules apply when you select the high throughput option:
- The vials must be arranged sequentially (for example, 1, 2, 3, and so forth).
- All instrument methods must be the same within a sample set.
- The number of injections for each line should be 1.
- Do not select "Alter running sample set".
When you select low throughput in the instrument method, you can perform the following operations:
- The vials can be arranged in a non-sequential manner.
- Instrument methods can be different within a sample set.
- The number of injections for each line can be more than 1.
- You can select "Alter running sample set".
ADDITIONAL INFORMATION
Applies to Agilent ICF 3.1 and later.
Agilent ICF Support Layer 3.1 Release Notes page 6.
id86822, HS7697B