What are the headspace rules that apply for high throughput and low throughput mode under the Options tab? - WKB86822

Article number: 86822

ENVIRONMENT

  • Agilent Instrument Control Framework
  • Agilent 7697A Headspace
  • Agilent G1888 Headspace
  • Empower 3 Enterprise
  • Headspace

ANSWER

When running samples, the following headspace rules apply when you select the high throughput option:

  • The vials must be arranged sequentially (for example, 1, 2, 3, and so forth).
  • All instrument methods must be the same within a sample set.
  • The number of injections for each line should be 1.
  • Do not select "Alter running sample set".

When you select low throughput in the instrument method, you can perform the following operations:

  • The vials can be arranged in a non-sequential manner.
  • Instrument methods can be different within a sample set.
  • The number of injections for each line can be more than 1.
  • You can select "Alter running sample set".

ADDITIONAL INFORMATION

Applies to Agilent ICF 3.1 and later.

Agilent ICF Support Layer 3.1 Release Notes page 6.