Empower calibration curve is missing a point - WKB86605

Article number: 86605

SYMPTOMS

  • A point is missing from the calibration curve; the rest of the points are present
  • Sample type is labeled as standard

ENVIRONMENT

  • Empower 3 Feature Release 4

CAUSE

Component editor is missing the amount/concentration.

FIX or WORKAROUND

  1. Go to Alter Sample. 
  2. Click on Edit Amounts.
  3. Specify the correct value in the component editor.
  4. Save Alter Sample.
  5. Reprocess the standards.