"Procsrvr MFC stopped working" error when processing data with internal standard - WKB86378
Article number: 86378
SYMPTOMS
- Batch processing a sample set with a processing method that includes an internal standard causes "Procsrvr MFC stopped working" error
- Batch processing the same sample set with a processing method with no internal standard works fine
ENVIRONMENT
- Empower 3 FR4
CAUSE
The method set included a processing method for a specific channel, but the internal standard in that processing method was from a different channel than that being processed.

PM "FGI" included a component from a channel called "QDa 3: SIR Ch1".
FIX or WORKAROUND
- Don't specify the processing method for a specific channel. Instead, process all channels in the method set using the same processing method.