Error "Fatal error. Empower will exit" when opening run sample in Empower 3 - WKB8531

Article number: 8531

SYMPTOMS

  • The following error occurs when opening run sample from Empower Pro interface: "Fatal error. Empower will exit."
    clipboard_ea92680c75390739d8d74e544cbd3afbe.png
  • The following error occurs when opening run sample from Empower project: "Failed to start application (Run Samples)"
    clipboard_e29905b31edaf41fe3908baf4829f6176.png
  • This appears for only one user with all projects and all chromatographic systems. Copy of Empower preferences does not resolve issue.

ENVIRONMENT

  • Empower 3 Enterprise
  • Citrix

CAUSE

Windows profile corruption.

FIX or WORKAROUND

  1. Delete the Windows profile of the user impacted by the issue on all Citrix servers.

ADDITIONAL INFORMATION

 