Slow processing speed in Empower when processing a sample set containing summarize custom fields function - WKB85206
Article number: 85206
SYMPTOMS
- A sample set is slow to process when summarizing custom fields
- The slow processing occurs only when processing a sample set containing impurities
- The project contains more than 20 custom fields, mostly intersample custom calculations
- Slow processing in Citrix and on an Empower workstation
ENVIRONMENT
- Empower 3
CAUSE
Project has multiple custom fields trying to calculate and summarize at the end of the sample set for multiple peaks.
FIX or WORKAROUND
In this instance, when either removing or modifying custom fields, the processing time decreases from 12 minutes to 2 minutes.
ADDITIONAL INFORMATION
This is normal behavior. Longer processing times occur with multiple custom fields and multiple peaks.