With Empower 3, empty Result Set for TUV data - WKB84410
Article number: 84410
SYMPTOMS
- "Process and report" selected in the Sample Set
- No data is present in the Result Set
ENVIRONMENT
- ACQUITY UPLC Tunable UV Detector
- Empower 3 Enterprise
- Empower 3 Personal
CAUSE
No processing method was selected in the Method Set used to process.
FIX or WORKAROUND
Add the appropriate processing method and select TUV Channel in the Method Set.