Skip to main content
Waters

How to change the unit for Amount in Empower 3 ? - WKB82956

Article number: 82956

OBJECTIVE or GOAL

Change the unit for Amount in Empower 3 ?

ENVIRONMENT

  • Empower 3
  • Empower 2

PROCEDURE

  1. In the Project window, open the Sample Set tab, right click on the appropriate sample set and click Alter Sample
  2. From the “Edit” menu in the Alter Sample window, select “Amount”.
  3. Modify the Unit column in the Component Editor window.
    ※ Although “<” is displayed in the Review window, a “<” must be entered on the left side of the unit entered.
    clipboard_e2e4f6543b0bb2299f0d217fc25927a3b.png
  4. Click OK to close the Component Editor window.
  5. Select File> Save in the Alter Sample window. to save the entered contents.

ADDITIONAL INFORMATION

Use the above method if you are just changing the unit for a specific sample set.
If you want to use the changed units in the future, modify the unit in the Sample Set Method.
In that case, perform step 1 of the above procedure as following:
1. Open the "Method" tab in the Project window, right-click on the appropriate sample set method and select Open.
The procedures after step 2 are the same as above.

 

Not able to find a solution? Click here to request help.