Empower error: Cannot save a result produced by processing a partially collected chromatogram - WKB82781
Article number: 82781
SYMPTOMS
- Processing channel and then saving it
- Error "Cannot save a result produced by processing a partially collected chromatogram"
- Sample set runs to completion
- Message Center indicates channel is still acquiring, yet status shows Acquisition Finished
ENVIRONMENT
- Empower 3
CAUSE
The Windows "allow control" permissions on the Empower folder are not selected for all permissions.
FIX or WORKAROUND
- First check the tablespace quota for the project (see WKB3096).
- Adjust Windows permissions on the Empower folder (allow full control/modify/read and execute, write). To troubleshoot, first try Everyone Full Control and confirm that system has full control on the raw data share folder.
- Allow the Everyone account to have full control for the Launch and Activation Permission settings for the remote DCOM on the Empower server.