Empower error with Agilent 1200: System not ready - WKB8123
Article number: 8123
SYMPTOMS
- Quaternary pump G1311A displays "System not ready" status
- The ICF Diagnostic log (c:\programdata\agilent technologies\instrument control framework) shows the following:
- Fatal: G1315B: DE60555714- Starting analysis aborted via CAN
Fatal: G1329A: DE60556342- Starting analysis aborted via CAN
Fatal: G1316A: DE60555486- Starting analysis aborted via CAN - Instrument status for Sampler, Column Compartment, and DAD is ready, but Quaternary Pump shows not ready
- Fatal: G1315B: DE60555714- Starting analysis aborted via CAN
- Run Samples displays "Waiting for Injection"
ENVIRONMENT
- Agilent modules:
1) Quat. Pump: G1311A
Firmware: A.06.10[005]
2) Sampler: G1329A
Firmware: A.06.10[006]
3) Column Comp: G1316A
Firmware: A.06.10[004]
4) DAD: G1315B
Firmware: A.06.10[004] - Empower 3 SR1 Enterprise
CAUSE
Communication issue
FIX or WORKAROUND
- Try a new method.
- If that doesn't work, replace the CAN cable on the instrument.