Oracle error with Empower 3: ORA-02291 : Integrity constraint (Milleinum.FK_SystemGroup_Systemnode) Violated - Parent Key not found - WKB81133

Article number: 81133

SYMPTOMS

  • Changing access from owner to owner group world generated an Oracle error: ORA-02291: Integrity constraint (Milleinum.FK_SystemGroup_Systemnode) Violated - Parent Key not found

ENVIRONMENT

  • Empower 3

CAUSE

During the migration from a test database to a production database, the option to use System Objects was selected. The production database still had reference to the test node.

FIX or WORKAROUND

  1. Delete and re-create the node.
  2. You can now create the new system and set changes as required.

ADDITIONAL INFORMATION

 