How to enable and collect 3D data using Agilent G7117C and Empower 3 - WKB80987

Article number: 80987

OBJECTIVE or GOAL

Enable and collect 3D data using Agilent G7117C and Empower 3.

ENVIRONMENT

  • Empower 3 Feature Release 4 (FR4)
  • Agilent DAD G7117C
  • Agilent ICF 3.1

PROCEDURE

  1. In the Empower instrument method, click on DAD.
  2. Click Advanced > Under Spectrum > Store > All.
  3. Specify Range and Step.
  4. Save the method.
  5. Collect data.
  6. Process using a PDA processing method and applying a method set.

 

G7117C 3D.png

 

 

ADDITIONAL INFORMATION

 