Empower freezes when trying to add a new solvent to the solvent catalog in the Empower Acquity instrument method - WKB78780
Article number: 78780
SYMPTOMS
- Empower freezes when trying to add a new solvent to the solvent catalog in the Empower Acquity instrument method
- If the catalog is opened and a currently present solvent is selected, Empower works normally
- A new "solventlibrary.xml" file on the node in question does not help
- The solvent field may be blank or not save new solvents
- This happens only on a particular node; others do not exhibit the problem
ENVIRONMENT
- Empower 3
- ICS 1.71
CAUSE
Bad ICS installation
Or wrong parameters in "C:\Empower\Instruments\Settings\AcquitySettings.xml" in <System_NONE> section (the window is hidden)
-<System_NONE>
-<SolventLibraryForm>
<HideDisabledLibraryEntries>true</HideDisabledLibraryEntries>
<Window>279,274,643,394</Window>
</SolventLibraryForm>
</System_NONE>
FIX or WORKAROUND
- Delete the "C:\Empower\Instruments\Settings\AcquitySettings.xml" file. (It will be created when you re-launch the application.)*
- Or reinstall the ICS on the affected node.