Saturated signal at 2250mV on eSAT/IN - WKB76803

Article number: 76803

SYMPTOMS

  • Saturated signal at 2250mV on eSAT/IN 

ENVIRONMENT

  • Empower 3
  • e-SAT/IN Module

CAUSE

Signal out of eSAT/IN specifications. (Analytical input range −0.25 to 2.25 Volts)

FIX or WORKAROUND

  1. Use 0-1V analog output on the detector
  2. Change conversion factor on the detector to have maximum signal less than 2.25V

ADDITIONAL INFORMATION

eSAT/IN specification : https://www.waters.com/waters/support.htm?lid=10170903&type=USRM