Saturated signal at 2250mV on eSAT/IN - WKB76803
Article number: 76803
SYMPTOMS
- Saturated signal at 2250mV on eSAT/IN
ENVIRONMENT
- Empower 3
- e-SAT/IN Module
CAUSE
Signal out of eSAT/IN specifications. (Analytical input range −0.25 to 2.25 Volts)
FIX or WORKAROUND
- Use 0-1V analog output on the detector
- Change conversion factor on the detector to have maximum signal less than 2.25V
ADDITIONAL INFORMATION
eSAT/IN specification : https://help.waters.com/help/en/support/library-details.html?documentid=71500049404