Modified method cannot be saved; errors include "currently being edited by" - WKB75209

Article number: 75209

SYMPTOMS

  • Unable to "save as" a modified method because it's currently being edited by "user" or is blank, Method Properties shows Being edited by User XXX and enabled the "Clear" Option.

clipboard_ece4102d0ac9d8181c6672e55665b669b.png

  • Method "method name" is read-only because it is currently being edited by "user name or system"

ENVIRONMENT

  • Empower 3 FR3
  • Empower 3 FR4

CAUSE

The project free tablespace value is less than 10 MB.

 

FIX or WORKAROUND

Adjust the project quota value to set free tablespace above 10 MB.

Open Method Properties and click on "Clear" and press "OK".