Agilent 6850/6890/7890 GC with a G1888A Headspace with Status "Waiting for Prep Run" using Empower - WKB74043
Article number: 74043
SYMPTOMS
- When running an Instrument Method that has Gas Saver enabled on an Agilent GC with a G1888 Headspace, the status stays at "Waiting for Prep Run".
- The Sample Set continues if the Prep Run button on the front panel of the GC is manually pressed.
- This happens whether or not the "Prep Run on Manual Request" is selected or cleared in the Pre-Configuration Utility (under the Configuration Tab of the GC).
ENVIRONMENT
- Agilent 6850 GC or Agilent 6890 GC or Agilent 7890 GC with an Agilent G1888 Headspace
- Empower 3
- Agilent Instrument Control Framework 3.0 and above
CAUSE
"Prep Run on Manual Request" does not work for the Agilent G1888. The G1888 does not supply a prep signal.
FIX or WORKAROUND
- Enable Auto Prep to "ON" through the front panel of the GC.
- Set the Auto Prep option to ON if you intend to run with an Instrument Method that utilizes Gas Saver Mode, Splitless Injection, Pulsed Split/Splitless Injection, or Solvent Vent Injection. This should be enabled for all samplers except ALS and the default value is to be enabled for all samplers except ALS.
ADDITIONAL INFORMATION
CRI-946 was created for this defect
PCS 44338 There is a similar issue with "native" drivers.
See article WKB62833 on split injection technique.
See article WKB62831 on splitless injection technique.
See article WKB62933 on PTV injection technique.
See article WKB97285 on Headspace Sample preparation
id74043, 6890A/+GC, 6890N GC, 7890 GC, 7890 GCDET, 7890BGC, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, HEADSPACE, SUP