Instrument failure with Agilent 7890 or 6890 and Empower 3 - WKB73623
Article number: 73623
SYMPTOMS
- Instrument failure with Agilent 7890 or 6890
ENVIRONMENT
- Empower 3
- Agilent Instrument Control Framework Support v3.1
CAUSE
The Agilent GC cannot reach the temperature set in the instrument method within 10 minutes (for example, detector, inlet temperature for shutdown method)
FIX or WORKAROUND
Perform injections using some temperature steps (with the steps reachable within 10 minutes).
ADDITIONAL INFORMATION