Error "Processing Method Editor: Attempted to access an unnamed file past its end" when attempting to edit a processing method in Empower - WKB73450
Article number: 73450
SYMPTOMS
- When attempting to edit a processing method, the error "Processing Method Editor: Attempted to access an unnamed file past its end" appears in Empower
ENVIRONMENT
- Empower 3
CAUSE
Multiple sessions of the method are open.
FIX or WORKAROUND
- Highlight the processing method > right-click.
- Click on method properties.
- Is the "clear" icon available? If so, click it.
- Now reopen the processing method and edit again.
ADDITIONAL INFORMATION