"Configuration not available" error when uploading configuration with Agilent Preconfiguration using Empower - WKB72591
Article number: 72591
SYMPTOMS
- "Configuration not available" error occurs when uploading configuration with Agilent Preconfiguration and the preconfiguration fails
ENVIRONMENT
- Agilent 7890 GC
- Agilent 7697A Headspace
- Agilent 6890 GC
- Agilent G1888 Headspace
- Empower 3 Enterprise
- Agilent Instrument Control Framework 3.0
CAUSE
Unwanted digits were entered in the Instrument IP address on the Connection tab in the Configuration dialog box.
FIX or WORKAROUND
Enter the correct Instrument IP address.
(Correct) 192.168.0.XX
(Wrong) 192.168.000.XXX
ADDITIONAL INFORMATION
See article WKB62833 on Split injection technique
See article WKB62831 on Splitless injection technique
See article WKB62933 on PTV injection technique
id72591, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, HEADSPACE, SUP