Empower sample set fails to run - WKB72466

Article number: 72466

SYMPTOMS

  • Sample set fails to run
  • Software waits to start first sample
  • Software was recently upgraded
  • "Sample: Wait for trigger" seen in Empower

ENVIRONMENT

  • Empower 3

CAUSE

Communication issue

FIX or WORKAROUND

  1. Close Empower.
  2. Reboot the Empower PC.
  3. Log in to Windows but do not launch Empower.
  4. Power-cycle all UPLC components, starting with the sample manager. Power-on the QDa last.
  5. Wait for the QDa status light to be steady green or OFF.
  6. Open Empower.

ADDITIONAL INFORMATION

Reboot the hardware after upgrading the software.