Cannot process sample set in Empower - WKB71726
Article number: 71726
SYMPTOMS
- Sample set does not process in Empower
- New processing methods do not work
- The message center contains no errors
- When the sample set is processed, selecting View as > Result sets for the project triggers the error "no result sets found for specified item"
- The audit trail notes that tablespace is below the minimum level
ENVIRONMENT
- Empower 3
- UV or MS Data
CAUSE
Project tablespace is too low to process and generate results.
FIX or WORKAROUND
Increase the project tablespace.