SYMPTOMS

  • Sample set does not process in Empower
  • New processing methods do not work
  • The message center contains no errors
  • When the sample set is processed, selecting View as > Result sets for the project triggers the error "no result sets found for specified item"
  • The audit trail notes that tablespace is below the minimum level

ENVIRONMENT

  • Empower 3
  • UV or MS Data

CAUSE

Project tablespace is too low to process and generate results.

FIX or WORKAROUND

Increase the project tablespace.