"Waiting for Injection" error with Shimadzu LC-2010 systems - WKB69489

Article number: 69489

SYMPTOMS

  • Empower message center and acquisition logs show the following error messages:
  • "Autosampler pretreatment is paused. Analysis is waiting to start until instruments become ready"
  • "-5310; Auto sampler pretreatment failed to complete. Analysis stopped."
  • "-5311; Instrument Failure LC-2010@4"

 

ENVIRONMENT

  • Empower 3 
  • Shimadzu i-Series Driver for Waters Empower Software Version 1.00
  • Shimadzu LC Driver for Waters Empower Software Version 3.10 SP6
  • Shimadzu LC Driver for Waters Empower Software Version (Nexera/Prominence/LC-2010) Version 4.42

CAUSE


The oven did not stabilize because of fluctuations in room temperature. It took a long time to stabilize because of a malfunction in cooling of the LC-2010 oven. Because the door of the sampler was not closed firmly, vibration opened it during operation.

FIX or WORKAROUND

The Shimadzu LC system should be stabilized before starting a sample set.