With Empower, next injection does not start on the Agilent 7890B GC - WKB68297

Article number: 68297

SYMPTOMS

  • With Agilent GC 7890B, next injection does not start if instrument method has a thermal gradient
  • Without thermal gradient (isothermal), there is no issue

ENVIRONMENT

  • Empower 3 FR3
  • Agilent ICS 7890B V2.6

CAUSE

The GC 7890B has been used with other software (MSD Chem). This software changed some parameters on the GC 7890B.
Configuration of GC 7890B Oven : Ignore Ready was set to "True". 

FIX or WORKAROUND

Change GC 7890B Oven : Ignore Ready to "False" on the control panel of the GC 7890B.

ADDITIONAL INFORMATION

id68297, 7890 GC, 7890 GCDET, 7890BGC, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP