With Empower, next injection does not start on the Agilent 7890B GC - WKB68297

Article number: 68297

SYMPTOMS

  • With Agilent GC 7890B, next injection does not start if instrument method has a thermal gradient
  • Without thermal gradient (isothermal), there is no issue

ENVIRONMENT

  • Empower 3 FR3
  • Agilent ICS 7890B V2.6

CAUSE

The GC 7890B has been used with other software (MSD Chem). This software changed some parameters on the GC 7890B.
Configuration of GC 7890B Oven : Ignore Ready was set to "True". 

FIX or WORKAROUND

Change GC 7890B Oven : Ignore Ready to "False" on the control panel of the GC 7890B.