No calibration curve in Empower 3 - WKB6616

Article number: 6616

SYMPTOMS

  • No calibration curve generated with result set
  • Peak codes C01, C02, Q20

ENVIRONMENT

  • Empower 3

CAUSE

  • In Component Editor, the values in the columns for purity factor and label claim were swapped

FIX or WORKAROUND

Enter the appropriate information in Alter Samples and Component Editor.

ADDITIONAL INFORMATION

 