Amounts are not displayed in Empower - WKB6362

Article number: 6362

SYMPTOMS

  • Amounts are not calculated
  • Standard 1 peak not integrated

ENVIRONMENT

  • Empower 3 personal

CAUSE

Minimum height set too high in processing method

FIX or WORKAROUND

  1. Adjust minimum height using the lowest concentration standard.
  2. Reprocess data.

ADDITIONAL INFORMATION

The same can be said for the minimum area.