Skip to main content
Waters

Instrument failure occurs when logging out of Windows operating system - WKB61802

Article number: 61802

SYMPTOMS

  • "Instrument failure" occurs when logging out of Windows operating system
  • Acquisition data is not collected

ENVIRONMENT

  • Empower Personal workstation

CAUSE

Windows user-launched services, including Empower application services, are shut down when the user exits the operating system.

FIX or WORKAROUND

Do not log out of Windows during an active running sample set on an Empower Personal installation.

ADDITIONAL INFORMATION

Fast user switching can bring a similar result. See KCS WKB2017.

id61802, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP

Not able to find a solution? Click here to request help.