PDA overlay function is excluded from Empower - WKB61700

Article number: 61700

SYMPTOMS

  •  It is no longer possible to overlay extracted chromatograms from 3D data channels between injections

ENVIRONMENT

  • . Empower 3 Feature Release 4, Service Release 3

CAUSE

Functionality was removed to prevent crashes in Review due to lack of memory addressability.

FIX or WORKAROUND

Bring chromatograms into Review, extract, process, save, and then overlay from results.

ADDITIONAL INFORMATION

Refer to PCS 56487.

 