How to configure Agilent GC-HS communications - WKB58496

Article number: 58496

OBJECTIVE or GOAL

Configure Agilent GC-HS communications.

ENVIRONMENT

  • Empower 3
  • Agilent ICF 3.0, A.03.02 (GC driver)
  • Agilent 7890B GC
  • Agilent 7697A Headspace Sampler
  • Agilent G1888 Headspace Sampler

PROCEDURE

The headspace sampler can be configured to communicate with the connected GC. This feature requires a HS with firmware A.01.06 (or greater) and a compatible Agilent GC (for example, a 7890B GC with firmware B.02.02 or greater). With communications configured, the HS knows GC method timings and programs, synchronizes with the GC clock, and follows the GC instrument schedule.

To configure GC-HS communications:

  1. At the headspace sampler keyboard, press [Options], and then go to Communications > GC.
  2. Enter the GC's IP address, and then press [Enter].
  3. At the GC keyboard, press [Front Injector] if the headspace sampler is connected to the GC front inlet, or press [Back Injector] if it is connected to the GC back inlet.
  4. Scroll to No Headspace, and then press [Mode/Type]. A list of headspace sampler models appears. Scroll to the correct one, and then press [Enter].
  5. Press [Options], and then select Communications.
  6. Scroll to the IP address entry for the headspace sampler, either Fr. Headspace or Bk. Headspace. Enter the headspace sampler's IP address.
  7. Press [Front Injector] or [Back Injector] as applicable.
  8. A new Communication line appears. Scroll to it and press [On/Yes]. The line changes to read Connected time. Scroll down to display a read-only summary of the headspace sampler configuration that the GC obtained from the sampler.
  9. If communications fail, check the IP addresses. Verify that the GC's IP address is correct in the headspace sampler, and that the headspace sampler's IP address is correct in the GC. Also verify that both instruments are turned on and connected to the LAN. Verify all LAN cable connections.

To temporarily disable communications:

  1. At the GC keyboard, press [Front Injector] or [Back Injector] as applicable.
  2. Scroll to Connected time and press [Off/No].

This setting disables communications but leaves the configuration information intact.

ADDITIONAL INFORMATION

This information is found in the help menu for the Headspace Control via the Empower Configuration Manager Preconfiguration Utility under the Tools pull-down menu.

Not able to find a solution? Click here to request help.