Error: Unable to load instrument setup information for AgGC - WKB58222
Article number: 58222
SYMPTOMS
- Instrument method in report shows error logged as "Unable to load instrument setup information for AgGC"
- Empower message center shows error "Unable to load instrument setup information for AgGC"
ENVIRONMENT
- Empower 3
- ICF 2.2 or 3.0 with A.02.04 and A.02.05 ICF framework and GC driver A.03.02 or B.01.03a
- ICF SL v3.3
- 6890N
- Agilent Instrument Control Framework 7890 GC
CAUSE
System is not accepting "JW" as column manufacturer in instrument method.
FIX or WORKAROUND
- Reopen the instrument method.
- In the Column tab, click on Column Catalog.
- Change the Manufacturer from JW to Agilent.
- Save the method again.
ADDITIONAL INFORMATION
Defect ID: INFICFSL-205