Skip to main content
Waters

Using a sample set template in Empower, the populated plates are not sequential - WKB5815

Article number: 5815

SYMPTOMS

  • The sample set bypasses plate 2 when populating the sample set

ENVIRONMENT

  • Empower 3
  • Dissolution option
  • Sample set template
  • ACQUITY H-Class with Sample Organizer

CAUSE

The defined default plate configuration was out of sequence when the sample set template was created.

Incorrect plate configuration.JPG

FIX or WORKAROUND

  1. Correct the defined plate configuration: From the Run Samples window, select Customize > Defaults > Plates. Select Define, and then select the plate being used. (Tip: If using standard plates, select ANSI-48Vial2mLHolder for each plate position.) Select OK > OK.
  2. Select File > Save preferences.
  3. Create a new sample set template.

ADDITIONAL INFORMATION

 

Not able to find a solution? Click here to request help.