Cloning projects and subprojects is interrupted - WKB56905
Article number: 56905
SYMPTOMS
- Projects are cloned
- Cloning process aborts and does not continue
ENVIRONMENT
- Empower 3
CAUSE
A single project is set to "Full Lock" status. This aborts the cloning operation.
FIX or WORKAROUND
Change the project status to something other than "Full Lock", and copy to the destination.
ADDITIONAL INFORMATION
In this case, all projects listed after the "Full Lock" project needed to be manually selected and copied to the destination folder.