Empower report signoff fails because user is logged in from multiple nodes - WKB56892
Article number: 56892
SYMPTOMS
- Empower signoff fails, indicating that user is logged in from multiple locations
ENVIRONMENT
- Empower Enterprise
- Empower Workgroup
- Citrix
CAUSE
The Empower System Policy "Enforce Single Logon for Sign Off" is enabled.
The Empower user did not properly logout of the operating system/Empower on all nodes.
An Empower Citrix session moved from the original Citrix client node to a second Citrix client node. Empower treats this as multiple logons.
FIX or WORKAROUND
- Log out of Empower on all nodes. Log out of the the OS on all nodes. Log back in to Empower and re-attempt signoff.
or
2. Disable the Empower System Policy "Enforce Single Logon for Sign Off".
ADDITIONAL INFORMATION
This is a known issue , CRI-1524. This is currently under investigation.
See web posting for further details
CRI-1524 - Empower treats moving Citrix session as multiple logons : Waters