No comment required when I back up and then delete a project in Empower? - WKB5653
Article number: 5653
SYMPTOMS
- No comment required when backing up and deleting a Project in Empower
ENVIRONMENT
- Empower 3
CAUSE
In the Configuration Manager, under View > System Policies, on the system audit trail tab, project is set to "Silent"
FIX or WORKAROUND
- Go into the Empower Configure the System.
- Click View > System Policies.
- Select the System Audit Trail tab.
- In the comment section, choose "Unrestricted" to allow free text comments, or "Restricted" to limit the user to pre-defined comments.
- Click OK.