Empower error in Run Samples: System idle - Run aborted - Instrument already in use - WKB5578
Article number: 5578
SYMPTOMS
- Error message: System idle - Run aborted - Instrument already in use
ENVIRONMENT
- Empower 3
CAUSE
A process did not close properly
FIX or WORKAROUND
- Close Run Samples.
- Go to Configure the System, select Systems, and right-click the system name in the table.
- Select Take Offline.
4. Power-off all instruments.
5. Reboot the LAC/E or workstation.
6. Power-on all instruments.
7. Select Bring Online.