e-SAT/IN does not collect data during an injection in Empower - WKB51699

Article number: 51699

SYMPTOMS

  • e-SAT/IN does not collect data during an injection but can monitor baseline
  • Pressure trace for 2695 collects correctly during an injection as well as a data channel from 2996 PDA detector 

ENVIRONMENT

  • Alliance 2695
  • 2997 Photodiode Array Detector
  • e-SAT/IN
  • Empower
  • MassLynx
  • 2414 RI Detector

CAUSE

One of the ends of the inject start cable is not secured or not making contact on the e-SAT/IN. 

FIX or WORKAROUND

Re-seat and secure the inject start cable.

ADDITIONAL INFORMATION

 

id51699, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SATIN, SUP