e-SAT/IN does not collect data during an injection in Empower - WKB51699
Article number: 51699
SYMPTOMS
- e-SAT/IN does not collect data during an injection but can monitor baseline
- Pressure trace for 2695 collects correctly during an injection as well as a data channel from 2996 PDA detector
ENVIRONMENT
- Alliance 2695
- 2997 Photodiode Array Detector
- e-SAT/IN
- Empower
- MassLynx
- 2414 RI Detector
CAUSE
One of the ends of the inject start cable is not secured or not making contact on the e-SAT/IN.
FIX or WORKAROUND
Re-seat and secure the inject start cable.
ADDITIONAL INFORMATION
id51699, 1525, 1525U, 1525UIVD, 2414, 2414B, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, A-10RI, A-10SSM, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, E-SATIN, E-SATIN SW, MLYNX, MLYNXV41, SATIN, SUP, SUPMM