During processing in Empower, what kind of operation will generate a “Created Calibration” entry in audit trail? - WKB50288

Article number: 50288

ENVIRONMENT

  • Empower 3

  • Project full audit trail

ANSWER

The audit trail of "Created Calibration" will be displayed if “Clear Calibration” was activated during processing.

Even if there is no Standard in the sample set, when "Clear Calibration" is included in the first row of the sample set, the "Created Calibration" entry will be generated during processing.

When performing background processing, if the "Clear Calibration” check box is selected, the "Created Calibration" entry will also be generated.