With Empower 3 FR4 and Windows 10, DHCP is empty after Windows reboot - WKB50032
Article number: 50032
SYMPTOMS
- After shutdown of the PC, the WatersDHCPserver is empty
ENVIRONMENT
- Empower 3 FR4
CAUSE
The "Turn on fast startup (recommended)" option is enabled in the Windows power settings.
FIX or WORKAROUND
- Type "Control Panel" in Windows 10 search, and open the Control Panel.
- Click on the "Power Option" icon.
- On the left, click on "Choose what the power buttons do".
- Click on "Change settings that are currently unavailable" and clear the "Turn on fast startup (recommended)" selection.

ADDITIONAL INFORMATION