2489 "Not Connected" error in Empower Run Samples from another node - WKB48930
Article number: 48930
SYMPTOMS
- "Not Connected" error in Empower Run Samples control panel from another node (control panel ok from the LAC/E32 where the 2489 is connected)
- Reset and reboot does not resolve
- Node Properties shows status = Yes under OK?
- LED is green on the 2489
- Can ping the detector and get reply
- No issue if connected to Empower from the node where the 2489 is connected
ENVIRONMENT
- 2489 detector
- Ethernet
- Empower 3 Enterprise
CAUSE
IPV6
FIX or WORKAROUND
ADDITIONAL INFORMATION
