Why does a new Empower LAC/E32 installation buffer data and give the error "unable to connect to node" in the recover.log file? - WKB48704
Article number: 48704
SYMPTOMS
- You can log in and start a run on the new LAC/E32, but it buffers data and gives an "unable to connect to node" error in the recover.log file
ENVIRONMENT
- Empower 3
CAUSE
The tnsnames.ora file does not point to the correct Empower server database instance.
FIX or WORKAROUND
- Correct the tnsnames.ora file on the problem LAC/E32, or
- Copy a working tnsnames.ora file from a working LAC/E32 to the problem LAC/E32.
ADDITIONAL INFORMATION
You may need to restart the LAC/E32 to re-establish a good connection for the Watersservice to copy buffered raw data files to the raw data share on the Empower database server.