Where is the sensitivity setting for a 2414 RI detector under Empower control? - WKB46956
ENVIRONMENT
- 2414 Refractive Index (RI) Detector
- ACQUITY Arc Driver v1.65 with 2414 firmware v2.10
- Empower 3 Feature Release 4
- Breeze 2
- Empower QS
ANSWER
Sensitivity is no longer an available setting.
ADDITIONAL INFORMATION
Sensitivity was required only for 410 emulation mode.
410 emulation mode explanation:
In 410 mode, the detector monitors a single channel, in millivolts, and does not support standalone method programmability. Control is provided via the data system software or Alliance system. You can configure several additional parameters when using the detector in 410 mode: • Sensitivity – Used in the scaling of the refractive index data in millivolts and affecting the front panel display, analog outputs, and remote interface data. Sensitivity is used only to generate Waters 2410/410 RI Detector-compatible data for older data systems and controllers.
id46956, 2414, 2414B, A-10RI, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP