How to activate gradient ability for Shimadzu LC-20AT pump - WKB3435

Article number: 3435

OBJECTIVE or GOAL

Activate the gradient functionality for Shimadzu system with LC-20AT pump configured with optional LPGE (low-pressure gradient unit).

ENVIRONMENT

  • Shimadzu driver v4.02
  • Empower 3 FR3

PROCEDURE

  1. Access the Shimadzu Instrument configuration wizard.
  2. Select the correct IP address for CBM-20A/Lite and go to the second page.
  3. Specify properties button for the LC-20AT pump.
  4. Select LPGE Unit from drop-down menu for Solenoid valve.
  5. Complete wizard steps.
  6. Log in to Empower and edit the instrument method to use the gradient function instead of the default isocratic mode.

ADDITIONAL INFORMATION

 