Empower fails on launch/login - WKB330201
Article number: 330201
SYMPTOMS
- Application Empower.exe fails during launch or logon
- No UI or login screen appears, or failure occurs immediately after a login attempt on Empower workstation with v3.10.0 installed
- Windows Event Viewer shows:
- Faulting application name: Empower.exe
- Faulting module name: KERNELBASE.dll
- Exception code: 0xebf00bab (or 0xe06d7363)
ENVIRONMENT
- Empower 3
CAUSE
The failure occurs because Empower cannot start the required .NET runtime during initialization.
FIX or WORKAROUND
- The .NET Runtime (x86) required by Empower is missing or incompatible.
- Ensure that the .NET Runtime version is 8.011 (x86) and the .NET ASPNET Core Runtime version is 8.011 (x86).
- Use the Setup Prerequisites folder from the media to install the .NET Runtime and VC redistributable packages.