Oracle error ORA-02000: missing error during Project clone or copy - WKB32710

Article number: 32710

SYMPTOMS

  • Copy or clone project is unsuccessful
  • CONFIGMGR MFC error
  • ORA-02000 message

ENVIRONMENT

  • Empower

CAUSE

There is a an issue with custom field naming in Oracle. The problem is usually in custom fields of a calculating type (for example Real, Peak, or Result) that would contain a formula.

FIX or WORKAROUND

  1. Rename the custom field by replacing all spaces with underscores ("_").