Skip to main content
Waters

The diverter valve does not switch back to the waste position per the last line of the Events table in the Empower instrument method or MassLynx experiment setup- WKB32237

Article number: 32237

SYMPTOMS

  • Diverter valve goes to the correct position at start of method
  • Diverter valve changes to the designated position per the Event Table
  • Diverter valve does not change to the designated position per the last event in the Event Table

ENVIRONMENT

  • SQ Detector 2
  • Empower 3, SR1, FR1, SR2HF1, SR2, SR2HF1
  • SQD2 ICS v1.51.1644
  • Xevo TQ-S micro
  • MassLynx SCN945

CAUSE

The time for last event is in the event table is the same as the end time for the Empower instrument method or MassLynx experiment setup.

FIX or WORKAROUND

Change the event time for the last event to be at least 0.01 minutes less that the end time of the Empower instrument method or MassLynx experiment setup.

ADDITIONAL INFORMATION

 

Not able to find a solution? Click here to request help.