Skip to main content
Waters

Empower 3 instrument method "Save with Method Set" does not work properly - WKB31492

Article number: 31492

SYMPTOMS

  • When creating a new instrument method and selecting "Save with Method Set", the user is not allowed to provide a name for the instrument method
  • The name shown is "Untitled", and an instrument method named "Untitled" exists in the project

ENVIRONMENT

  • Empower 3

CAUSE

There is an instrument method in the project with the name "Untitled".

FIX or WORKAROUND

  1. It is best to delete the instrument method named "Untitled" from the project.
  2. Another option is to save the instrument method and the method set in separate actions, instead of using "Save with method Set".

ADDITIONAL INFORMATION

It is unusual to have an instrument method in a project with the name "Untitled"; users are not allowed to save to this name.

This issue has been observed with a German operating system, or possibly with regional settings.

id31492,

Not able to find a solution? Click here to request help.