Empower error: Cannot save a result produced by processing a partially collected Chromatogram - WKB3096
Article number: 3096
SYMPTOMS
- Reviewing a sample set that contained 532 injections
- Applying processing method and then saving results
- Get the error "Cannot save a result produced by processing a partially collected Chromatogram" when trying to save results
- Sample set ran to completion
ENVIRONMENT
- Empower 3
CAUSE
Low database tablespace for project
FIX or WORKAROUND
Increase the Quota for the Database Tablespace:
- Select Configure the System.
- Select Projects on the left-hand side.
- Highlight the project.
- Right-click and select "Properties".
- Change the value of the Database Tablespace Quota. (Example was at 100 MB with only 1 MB Free; increased Quota to 166 MB.