SYMPTOMS

  • Reviewing a sample set that contained 532 injections
  • Applying processing method and then saving results
  • Get the error "Cannot save a result produced by processing a partially collected Chromatogram" when trying to save results
  • Sample set ran to completion

ENVIRONMENT

  • Empower 3

CAUSE

Low database tablespace for project

FIX or WORKAROUND

Increase the Quota for the Database Tablespace:

  1. Select Configure the System.
  2. Select Projects on the left-hand side.
  3. Highlight the project.
  4. Right-click and select "Properties".
  5. Change the value of the Database Tablespace Quota. (Example was at 100 MB with only 1 MB Free; increased Quota to 166 MB.