Error: "Database failure storing injections" in Empower 3 - WKB3081
Article number: 3081
SYMPTOMS
- In run sample, cannot acquire data
- Error message: "Database failure storing injections"
- Error message: "Database failure storing chromatogram"
ENVIRONMENT
- Empower 3 feature release 2
CAUSE
Project quota is insufficient.
FIX or WORKAROUND
- Increase project quota.
ADDITIONAL INFORMATION
See KCS article WKB1486 for how to increase quota.
id3081, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP