PA800plus CE Empower 3 - GPIB-USB communication issue - WKB30503
Article number: 30503
SYMPTOMS
- No instrument connection; communication issue
- The instrument stays offline in Empower
ENVIRONMENT
- Beckman PA800plus
- Empower 3 FR4 (Workstation)
- National Instruments NI GPIB-USB-HS
CAUSE
National Instruments GPIB-USB-HS+ interface cable does not correspond to the listed driver (3.1.1).
FIX or WORKAROUND
Install an older version of interface cable NI GPIB-USB-HS (not GPIB-USB-HS+ without the ferrite ring - see attached photo).
Another possible solution: Install new driver corresponding to the interface (requires interface configuration). Contact Beckman for support.