Unable to create a wavelength event in the instrument method of an ACQUITY PDA in Empower - WKB3033

Article number: 3033

SYMPTOMS

  • Events tab no longer allows for wavelength or channel entries. See the following images.

Events as seen using v1.50

PDA Events 1.50.JPG

Events as seen using v1.60 & v1.65

PDA Events v1.60.JPG

ENVIRONMENT

  • ACQUITY PDA
  • Driver Pack 4, v1.60
  • Driver Pack 4 Service Release 1, v1.65
  • Empower 3 Enterprise

CAUSE

Unknown.

FIX or WORKAROUND

  1. Upgrade to Driver Pack 2016 Release 2; this defect was corrected in ICS v1.68, which was released in Driver Pack 2016 Release 2.

ADDITIONAL INFORMATION

The work-around is to collect multiple 2D channels or collect 3D data and extract each channel separately.